Axis 670 Bedienerhandbuch Seite 53

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 68
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 52
6.2. 6740-XX Notch FUter Module
6.2.1 Background Theorv
All mechanical systems are subject to vibrations via extemal excitation forces. The degrees of
freedom of a vibrating mechanical system are defined by the number of independent coordinates
required to identify its displacement during vibration. If we have a Cartesian coordinate system
witii
X,
y, and z axis, then for a freely vibrating body, we can have six degrees of
freedom.
This
includes translational and rotational vibration in each of the three axis. Each of these possible
vibrations is referred to as a mode of vibration. Each mode of vibration has a natural frequency
associated with it that is independent of all the other modes.
For our discussion with respect to the scanner, we will concenfrate on the rotational axial mode
of
vibration.
The natural frequency Fr of this mode is a fimction of the mirror load inertia and
rotor inertia as well as the torsional spring constant of the rotor shaft which couples the two
inertias. The undamped natural resonant frequency of this mechanical system is described by the
equation below:
COR
= k'^[(J,+J2)/J,J2]''', Fr=(OR/27t
where k = rotor torsional spring constant
Jl = mirror load inertia
J2
= rotor inertia
Note that in a real mechanical scanner system damping due to bearing friction, air friction on the
rotor and mirror load, etc. do indeed exist. However, CTI scanners exhibit very low bearing
friction, and the above equation will approximate the resonant frequency quite closely. The
damping constant inversely influences the natural resonant
frequency
slightly. In other words, as
the damping constant increases, the natural resonant
frequency
decreases.
This torsional resonant frequency can exhibit a high Q, which is defined in one sense as the
sharpness and amplitude of the resonant frequency peak. If this resonance occurs in a closed
loop servo system where the gain vs
frequency
has fallen off enough that the (negative) position
feedback phase shift approaches
180°,
and the amplitude peaking at resonance is enough that the
position feedback gain rises near unity, servo loop instability and oscillations will result. This is
the reason for using a notch filter in the forward path of the servo loop, after the summing
amplifier. The notch filter is tuned to remove the
frequency
component ofthe ertor drive signal
sent to the scanner confrol coil which is at the same frequency as the scanner torsional
resonance. This keeps the scanner from being excited at its resonant
frequency.
The rejection of
this driving
frequency
aids the stability ofthe servo by not exciting this natural resonant mode,
and allows the closed loop bandwidth ofthe system to be higher. Higher closed loop bandwidth
allows decreased step response times.
However, this does not mean that the scanner torsional resonance dissappears. Ideally, it is only
that the servo amplifier no longer
"kicks"
the scanner at this frequency.
53
Seitenansicht 52
1 2 ... 48 49 50 51 52 53 54 55 56 57 58 ... 67 68

Kommentare zu diesen Handbüchern

Keine Kommentare