6.1.5.
Adjusting the Position Output Scale Factor and the AGC Linearity
The Position Output Scale Factor is precisely adjusted at Cambridge Technology and under
normal circumstances never needs adjusting by the customer for the life of the servo board or
scanner. If however the customer has changed the scanner originally sent with the servo board,
(remember they are matched sets) use the following procedure to verify/adjust this signal.
The most important signal generated on the 670 board is the Position Output, Vp, signal. In
order for the servo and the error protection circuitry to work correctly, the Vp signal must be
scaled properly. The Position Output Scale Factor is conttolled by increasing or decreasing the
AGC voltage sent to the scanner's position detector. The Position Output Scale Factor is linearly
proportional to this AGC voltage. Thus, changes in the Position Output Scale Factor causes
changes in the amount the scanner will move for a given output response from Vp. R13 is used
to adjust this AGC signal. This trimpot allows-10% adjustment range. If
the
AGC voltage is
changed by more than
1%
during this adjustment, the tuning ofthe system must be verified to be
sure that it is still set properly. It is always pmdent to check the tuning after any adjustment to
the Position Output Scale Factor or the Linearity Adjustment.
The Linearity Adjustment is used to improve the linearity of the position detector and position
demodulator. By varying the amount that Va and Vb sums together to create the AGC voltage,
the linearity is improved. The intention of
this
adjustment is to minimize the amount the AGC
voltage varies as a function of the angular position of the scanner. Since the Linearity
Adjustment and the Position Output Scale Factor adjustments affect each other, they must both
be set at the same time.
Note: Do not allow the AGC signal present at TP7 to exceed 11 volts or the AGC circuit may
saturate, which would result in improper operation.
••Caution: In all cases, the position signal, Vp, must always have the ability to exceed +/-10
volts when the scanner shaft is tumed to its intemal mechanical stops or scanner damage may
occur.
6.1.5.1
Closed Loop Method
If the system is afready tuned up and verification or "touching up" of the Position Output Scale
Factor and Linearity Adjustment are desired, use of this method is allowed. Also, if mixing and
matching scanners, this method can be used. A mirror must be mounted on the scanner shaft, or
some other means of measuring the scanner's angular position must be employed. By reflecting a
laser beam from the mirror to a wall, and using some simple trigonometry, the Position Output
Scale Factor can be adjusted with high resolution. The further from the wall, the more accurate
this method.
1.
Follow steps
1
- 7 in section 4.2. above. Set up the input for analog single ended operation
by setting Jl and W4 as indicated in section 3.6.
28