ECD Detector
Item Description
Detector Oven On or off.
Turns the TCD oven on or off.
Electronics On or Off.
Indicates whether the installed ECD will be
used or not.
Temperature (C) 50- 450 C.
Time Initial - 999.99 min.
Range 1 or 10.
Autozero Yes or No.
Set to yes at initial time means that the
ECD Autozero is on continuous before the
run starts. Autozero is automatically
disabled once the run starts.
Add Adds a line to the spreadsheet.
Insert Inserts a line above the currently selected
row in the spreadsheet.
Delete Deletes the currently selected row(s) in the
spreadsheet.
Sort Sorts the spreadsheet rows by time.
Time Constant Slow or Fast.
Cell Current N2 High, N2 Std, CAP, Ar-CH4, Zero.
Contact Potential (mV) 0- 800 mV.
EFC Type None, Type 13 for the ECD.
Make up Flow (ml/min) 0-50 ml/min.
Methanizer Installed Yes or No.