Axis BTA 2100D Betreibershandbuch Seite 258

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 597
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 257
256
3800 GC Column Oven
Item Description
Column Oven Coolant On or Off.
Specifies whether or not the column coolant will be used.
Enable Coolant at (C) 30- 450 ºC
Specifies the temperature at which to enable the coolant.
Coolant Timeout (min) 0.01-999.99 min
Provides a safety measure to save coolant after the
specified time, if either the injector fails to reach its set
temperature or the GC does not go into run.
Stabilization Time (min) 10.0 min
Specifies the column Stabilization Time.
Temp (C) -99 - 450 ºC
Rate (C/min) 0.01- 100.0 ºC/min
The Rate in the first row is always blank and cannot be
edited.
Hold (min) 0.01-999.99 min
Total (min) 0.01-999.99 min
Cannot be edited. Column Oven End Time (Total) is
displayed in all time-programmed windows of the Method.
Insert Inserts a line above the currently selected row in the
spreadsheet.
Delete Deletes the currently selected row(s) in the spreadsheet.
Seitenansicht 257
1 2 ... 253 254 255 256 257 258 259 260 261 262 263 ... 596 597

Kommentare zu diesen Handbüchern

Keine Kommentare