ADT Model 7100 Semi-Automatic Dicing System Maintenance Manual
Quarterly Maintenance Procedures (page 1 of 3)
P/N 97100-9005-000-14 Ver 06/05
3.3-1
3.3 Quarterly Maintenance Procedures
(page 1 of 3)
The maintenance procedures in this list should be performed on a
quarterly basis after performing the daily and monthly maintenance
procedures. The list includes the procedures to be performed, the
specifications that must be met, and a column for indicating whether the
System has passed each procedure.
Photocopy this list and mark each completed procedure.
Maintenance Technician Name: ________________________
Date: ________________________
Caution: All System operations must be stopped and any Workpieces
removed, before performing these procedures.
Quarterly Maintenance Procedures
Examine Procedure & Specifications Passed
Perform Daily
Maintenance
Procedures
See section 3.1.
Perform Monthly
Maintenance
Procedures
See section 3.2.
Remove old grease
from X, Y and Z Axes
See section 3.4.
Operation of the Main
Circuit Breaker switch
and the Emergency
Stop button
• While the System is powered on,
turn off the Main Circuit Breaker at
the right side of the System. Verify
that System operation stops
immediately.
• While the System is operating,
push the Emergency Stop button.
Verify that System operation stops
immediately.
Level of air pressure in
the System
Read the Air Pressure Gauge on the
Pneumatic Panel (see Figure 10-2). If
the air pressure is below 4-5atm, adjust
the Regulator located on the left behind
the Rear Side Lower Left Panel by
pulling the ring up, rotating the dial and
pushing the ring back into place.